Lee, Song, Won, Park, Chung, Lim, Ahn, and Lee: Comparing Inhaler Use Technique Based on Inhaler Type in Elderly Patients with Respiratory Disease

Abstract

Background

The aim of this study was to investigate inhaler device handling in elderly patients. Inhaler devices with respect to misuse and error correction were also compared.

Methods

Inhaler use technique was assessed using standardized checklists at the first visit and 3-month follow-up visit after retraining. The primary outcome was difference in the acceptable use ratio among inhaler devices. Secondary outcomes included differences in error correction, the most common step of misuse, and factors affecting the accuracy of inhaler use.

Results

A total of 251 patients (mean age, 76.4 years) were included. The handling of 320 devices was assessed in the study. All patients had been trained before. However, only 24.7% of them used inhalers correctly. Proportions of acceptable use for Evohaler, Respimat, Turbuhaler, Ellipta, and Breezhaler/Handihaler were 38.7%, 50.0%, 61.4%, 60.8%, and 43.2%, respectively (p=0.026). At the second visit, the acceptable use ratio had increased. There were no significant differences among inhaler types (Evohaler, 63.9%; Respimat, 86.1%; Turbuhaler, 74.3%; Ellipta, 64.6%; and Breezhaler/Handihaler, 65.3% [p=0.129]). In multivariate analysis, body mass index, Turbuhaler, and Ellipta showed positive correlations with acceptable use of inhalers, whereas Chronic Obstructive Pulmonary Disease Assessment Test score showed a negative correlation.

Conclusion

Although new inhalers have been developed, the accuracy of inhaler use remains low. Elderly patients showed more errors when using pressurized metered-dose inhalers than using dry powder inhalers and soft-mist inhalers. However, there were no significant differences in misuse among inhaler devices after individual training. Results of this study suggests that repeat training is more important than inhaler type.

Introduction

Inhaled therapy is the mainstay of treatment of airway diseases, such as asthma [1] chronic obstructive pulmonary disease (COPD) and tuberculosis-destroyed lung with airflow limitation [2,3], owing to the rapid onset of action and occurrence of fewer adverse events. Since the launch of inhalers in the 1960s, more than 250 inhaler devices have become available for effective treatment [4]. Following the introduction of pressurized metered-dose inhalers (pMDIs), several dry powder inhalers (DPI) and soft-mist inhalers were released, with differences in the spraying method, particle size, and device manipulation method [5]. However, the misuse of inhalers has been a longstanding discussion and is still a major concern for physicians [6-8]. Although decades have passed since the first inhaler was developed, the accuracy of inhaler use remains low at approximately 40% [8,9]. Poor inhaler technique is associated with poor disease control and frequent disease exacerbation [9,10].
The difficulty with correct inhaler use is more severe in elderly patients. Several studies revealed a significant positive correlation between the misuse of inhalers and age [11-13]. As cognitive function, physical strength and dexterity, memory, and executive planning deteriorate, errors in inhaler use increase in elderly patients.
Differences in the accuracy of inhaler use based on the inhaler type have been investigated in many studies. Two review studies were aimed at investigating whether the misuse differed with the type of inhaler or not but failed to elucidate which inhaler is more frequently misused [14,15]. However, some studies reported that the recently launched inhaler Ellipta had fewer errors of misuse compared to other inhalers [16,17].
The aim of this study was to evaluate how accurately inhalers are used by the elderly patients with respiratory disease and whether the usage accuracy varies with the inhaler type or not. In addition, we compared the inhaler devices with respect to error correction after training.

Materials and Methods

1. Study design

We conducted a prospective study to compare the accuracy of inhaler use among different inhaler types in elderly patients with respiratory disease at an outpatient clinic of the Veterans Health Service Medical Center, Seoul, South Korea, from June 2018 to July 2019. The accuracy of inhaler use was checked by a research investigator. If the patient made errors, the investigator demonstrated the correct inhaler use technique through a face-to-face discussion. Three trained research investigators checked the accuracy of the inhaler use technique with standardized checklists. We used a teach-to-goal approach to correct identified errors, giving feedback to the patient. We repeated this coaching process as many times as needed in order for patients to use inhaler correctly. After education, the accuracy of inhaler use was re-checked at the 3-month follow-up visit.
The study protocol was approved by the institutional review board of the Veterans Health Service Medical Center, Seoul (2018-04-035-003). This study complied with the principles laid down in the Declaration of Helsinki. Written informed consent was obtained from each patient before enrollment in the study.

2. Participants

Patients with asthma or COPD, aged 65 years or older, who had been using an inhaler for at least 3 months in addition to receiving inhaler training previously were enrolled in the study. Patients provided informed consent to participate in this study. We checked the inhaler use accuracy at the first and 3-month follow-up visits. COPD Assessment Test (CAT) scores, total duration of inhaler use, smoking history, education level, satisfaction, etc. were recorded. Spirometric analyses were performed at the time of inclusion in the study and 1 year later.

3. Checklists

We created a simple standardized checklist for this study. Regardless of the type of device, all checklists consisted of five steps. Steps 1 and 2 were to perform exhalation while assembling and holding the inhaler correctly. These steps varied from device to device. Step 3 was to breathing out fully. Step 4 was to perform deep inhalation. Step 5 was to hold the breath. Participants received a score of “0” for each incorrect step and a score of “1” for each correct step. The inhaler use was regarded “acceptable” if the patient performed four or more of the five steps correctly.

4. Outcomes

We compared five types of inhalers. The primary outcome was differences in the acceptable use ratio among inhaler devices. Secondary outcomes included differences in error correction after training, most common step of misuse, and factors affecting the accuracy of inhaler use. Further, we surveyed patient satisfaction with each device and found the most frequently error-prone steps. In this study, six inhaler device types were used by participants: Evohaler (GlaxoSmithKline, London, UK/Chiesi, Parma, Italy), Respimat (Boehringer Ingelheim, Ingelheim am Rhein, Germany), Turbuhaler (AstraZeneca, Lund, Sweden), Ellipta (GlaxoSmithKline), Breezhaler (Novartis, Basel, Switzerland), and Handihaler (Boehringer Ingelheim). However, Breezhaler and Handihaler were analyzed in the same group because of the similarities. Finally, five device types were compared.

5. Statistical analyses

Data on clinical characteristics are expressed as the mean±standard deviation or number (proportion). We used Fisher exact test and Kruskal-Wallis test, as appropriate, to compare baseline characteristics among groups. To identify the factors affecting acceptable use, we calculated the adjusted odds ratio and 95% confidence intervals using a logistic regression model. All analyses were performed using the STATA software, version 15 (StataCorp LP, College Station, TX, USA). Statistical significance was defined as a two-tailed p-value <0.05.

Results

A total of 251 patients were enrolled in the study. Table 1 summarized the baseline characteristics of the study participants. Patients included 234 men (93.2%), with a mean age of 76.38±0.39 years. The mean period of current inhaler device use was 37.60±2.83 months, and mean total period of inhaler use was 57.15±3.54 months. Two or more inhalers were used by 68 participants, so we assessed 320 devices in the study. Evohaler, Respimat, Turbuhaler, Ellipta, and Breezhaler/Handihaler were used by 75 (23.4%), 46 (14.4%), 45 (14.1%), 74 (23.1%), and 80 (25.0%) patients, respectively. At the first assessment, only 79 patients (24.7%) used the inhaler correctly.

1. Rate of acceptable users based on the inhaler device type

We defined “acceptable” if the patient performed four or more of the five steps in the checklist correctly. At the first assessment, there were 158 acceptable users (49.4%). There was a significant difference in the device use technique among the five types. Proportions of acceptable users for Evohaler, Respimat, Turbuhaler, Ellipta, and Breezhaler/Handihaler were 38.7%, 50.0%, 61.4%, 60.8%, and 43.2%, respectively (p=0.026) (Table 2, Figure 1). The mean checklist score (total of 5 points) was highest in Turbuhaler (3.71±1.16), followed by Ellipta (3.59±1.13), Respimat (3.46±1.38), Breezhaler/Handihaler (3.45±1.05), and Evohaler (2.89±1.54) (p=0.014) (Table 2).

2. Rate of acceptable users based on the inhaler device type among patients aged over 75 years

We analyzed inhaler use in 119 patients aged over 75 years. Proportions of acceptable users for Evohaler, Respimat, Turbuhaler, Ellipta, and Breezhaler/Handihaler were 37.0%, 42.9%, 62.5%, 51.7%, and 51.5%, respectively (p=0.547). The accuracy of inhaler use based on the inhaler device type among patients aged over 75 years was similar to the overall results, but no statistical difference was found.

3. Differences in the acceptable user ratio of the inhaler devices after individual education

When errors were corrected through one-to-one training and re-evaluated 3 months later, there were no differences in the rate of acceptable users among inhaler device types. Acceptable users for Evohaler, Respimat, Turbuhaler, Ellipta, and Breezhaler/Handihaler were 40 (64.5%), 31 (81.6%), 29 (76.3%), 43 (66.2%), and 49 (65.3%), respectively (p=0.280) (Table 3, Figure 1). The mean checklist scores (total of 5 points) also increased in all groups after training, but there was no difference between devices (Turbuhaler [4.13±0.84], Respimat [4.11±1.03], Ellipta [3.86±0.97], Breezhaler/Handihaler [3.84±1.12], and Evohaler [3.79±1.19], p=0.423) (Table 3).

4. Multivariate analysis of factors affecting inhaler use

Table 4 shows the results of univariate and multivariate logistic regression analyses of factors associated with the acceptable inhaler use in the baseline assessment. In the univariate analysis, the inhaler type, CAT score, body mass index (BMI), and the educational level were significantly associated with the acceptable use. No significant association with inhaler use was observed for patient age, presence of comorbidities, or period of use. In the multivariate analysis, the CAT score showed a negative correlation with the acceptable inhaler use whereas BMI, Turbuhaler, and Ellipta showed a positive correlation.

5. Comparison of the correct use rates at each step (based on the inhaler device) at baseline and after training

We evaluated the rate of correct handling of inhaler devices at each step at baseline and after training. Table 5 shows the results at the first visit and at the 3-month follow-up after training. Regardless of the type of inhaler device or training, patients performed “breathing out fully” least correctly. At the first visit, the percentage of patients who correctly performed the assembling step (pre-inhalation) was 64%-100%, and the patients who used Evohaler had the most errors (Evohaler, 64.9%; Respimat, 91.3%; Turbuhaler, 71.7%; Ellipta, 100%; and Breezhaler/Handihaler, 100%; p<0.001). For the fourth step of evaluating effective inhalation, which meant both full inhalation and coordination between inhalation and actuation, patients with Evohaler had the most errors (48% for Evohaler, 60.9% for Respimat, 93.3% for Turbuhaler, 82.4% for Ellipta, and 82.5% for Breezhaler/Handihaler; p<0.001). At the second visit after fixing the errors individually, the overall error rate had improved. However, errors by Evohaler users were significantly high in the assembling and inhalation steps.

Discussion

This was a prospective study performed to evaluate inhaler device usage, including newly developed devices, among elderly patients in the real world. Only 79 patients (24.7%) were used inhalers correctly, regardless of the inhaler device type. In a systematic review of errors in inhaler use from 1975 to 2014, Sanchis et al. [8] reported that the overall prevalence of the correct technique was only 31% (28%-35%) and has not improved over the past 40 years. Although new devices have been developed, these data are consistent with previous studies and demonstrate that the accuracy of inhaler use remains low.
We expected elderly patients to use DPIs better than pMDI owing to fewer coordination problems. In our study, Evohaler showed the lowest rate of acceptable inhaler technique among the five device types, and Turbuhaler and Ellipta showed fewer errors than MDI, with a statistical significance (p=0.026). In contrast, Respimat was not significantly different from DPI or MDI. After education, the significant differences in error frequency among devices disappeared. However, the analysis of each step revealed that Evohaler users performed more errors than users of other devices in the assembly (preinhalation) and inhalation steps even after training. Notably, Ellipta showed the lowest correction rate and Respimat showed the highest correction rate after one-on-one face-to-face training. We guess there are two main reasons. One is the selection bias arising from the design of this study. As a real-world study, we kept the device patients were using without randomly assigning devices to patients. Because Ellipta is easy to operate, there might be more patients with reduced cognitive and manipulation skills. The other reason for the lack of improvement is related to the stage at which the error appears. In the group of Ellipta, there were more errors in the exhalation (step 3), inhalation (step 4), and breath-hold (step 5) processes than in the preparation of device. In this study, the error of preparation step was relatively easy to be corrected, whereas the error of step 3-5 is difficult to improve. We thought that Respimat would require more individualized training, and if well-educated, errors could be reduced to a level over DPIs.
Although several studies showed that older age is a risk factor for incorrect use [13], the association between age and error is controversial [18,19]. In this study including the elderly people aged 65 years or above, age did not affect the accuracy of inhaler use in the multivariate analysis. Additionally, there was no difference in inhaler misuse between patients aged 65-75 and those aged 75 years or above. Diseases such as stroke or other factors that affect cognitive and behavioral skills, rather than age, may have a more significant impact on inhaler use [20]. Based on our results, we suggest that elderly patients can use inhalers suitably through repeated training.
Inhaler misuse is reportedly related to the device type (pMDI), female sex, lower education level, and lower socioeconomic status [21,22]. It is generally accepted that pMDI is more error prone than DPI, but studies are inconsistent in the findings of comparisons of DPIs, making it difficult to determine that a particular device shows fewer errors [22-24].
Many studies showed that educational intervention improves the accuracy of inhaler use. Klijn et al. [25] performed a systemic review of all studies related to education of inhaler technique and reported improvement in inhaler use, at least in the short term, with an average follow-up of 5 months. The teaching method, duration of education, individual, or group had no significant impact on outcomes [25]. Additionally, there was no difference in most variables correlated with poor inhalation between patients using pMDIs and those using DPIs [26]. The effectiveness of education also reportedly decreases over time [27,28]. Thus, it is recommended to train patients in inhaler use frequently, instead of switching inhalers [25]. The inhaler technique is unstable, so training on inhaler device use should be repeated to maintain correct use [18].
In this study, the most error-prone step was the third step of breathing out fully (correct users ratio, 46% [42%-50%]). This is a necessary step in inhaler use, regardless of the device type. As reported in previous studies, sufficient training is required for this step as well.
We assessed the accuracy of inhaler use in elderly patients, including new devices, and evaluated the degree of improvement after training and factors affecting inhaler use errors. However, there were limitations to the study. First, the patients kept the inhaler devices they were using and had not been randomly assigned to them. Therefore, it can be considered that the reality was reflected as it was, but our data cannot define an “easy-to-use inhaler.” The doctor would have already considered several factors related to the patient when choosing an inhaler. Physicians consider the ability to handle the device (97.9%), together with coordination (96.9%), ability to learn (87.5%), and adherence to therapy (84.4%) when selecting an inhaler device, based on the characteristics of each patient [29]. Second, we did not evaluate the patient’s cognitive or motor skills. We generally replaced these by checking the degree of education and history of stroke. Third, we could not measure the inspiratory flow and subjectively assessed the appropriate inhalation.
In conclusion, new inhalers have been developed, but the accuracy of inhaler use remains low in the elderly patients. The use of pMDI was associated with more errors compared to that of DPI or soft-mist inhalers, but there were no differences in misuse among inhaler devices after individual training. This study provides evidence that elderly patients need repeated training to use inhalers properly.

Notes

Authors’ Contributions

Conceptualization: Lee BJ. Methodology: Lee HY, Lee BJ, Chung KB. Formal analysis: Lee HY. Data curation: Lee HY, Won HK, Park Y, Chung KB, Lim HJ, Ahn YM. Software: Song JH, Chung KB. Investigation: Lee HY, Lee BJ. Writing - original draft preparation: Lee HY. Writing - review and editing: Lee BJ, Song JH, Won HK, Park Y, Chung KB, Lim HJ, Ahn YM. Approval of final manuscript: all authors.

Conflicts of Interest

No potential conflict of interest relevant to this article was reported.

Funding

This study was supported by grants from VHS Veterans Medical Research Institute.

Figure 1.
Proportions of acceptable use at the first visit and the second visit after individual education. The acceptable use rate by device at the first visit was 37.3% for Evohaler, 47.8% for Respimat, 62.2% for Turbuhaler, 60.8% for Ellipta, and 43.8% for Breezhaler/Handihaler (p=0.026). At the second visit after one-to-one training, there was no difference in the number of acceptable users for each inhaler type: 40 (64.5%) for Evohaler, 31 (81.6%) for Respimat, 29 (76.3%) for Turbuhaler, 43 (66.2%) for Ellipta, and 49 (65.3%) for Breezhaler/Handihaler (p=0.280).
trd-2020-0021f1.jpg
Table 1.
Baseline characteristics of patients
Characteristic Evohaler (n=75) Respimat (n=46) Turbuhaler (n=45) Ellipta (n=74) Breezhaler/Handihaler (n=80) p-value
Age, yr 76.2±6.1 75.6±6.4 74.9±5.4 76.5±6.1 76.8±6.3 0.547
Male sex 72 (96.0) 44 (95.7) 40 (88.9) 68 (91.9) 77 (96.3) 0.382
BMI, kg/m2 23.9±3.9 22.4±3.9 23.7±3.2 24.4±3.7 23.4±3.0 0.078
Smoking status 0.076
  Current smoker 11 (14.7) 12 (26.1) 10 (22.2) 14 (18.9) 17 (21.3)
Ex-smoker 59 (78.7) 33 (71.7) 26 (57.8) 51 (68.9) 58 (72.5)
Non-smoker 5 (6.7) 1 (2.2) 9 (20.0) 9 (12.2) 5 (6.3)
Education level 0.841
Lower (<6 yr) 32 (44.4) 16 (36.4) 13 (30.2) 27 (37.0) 34 (43.6)
Middle (6-12 yr) 33 (45.8) 23 (52.3) 23 (53.5) 36 (49.3) 37 (47.4)
High (>12 yr) 7 (9.7) 5 (11.4) 7 (16.3) 10 (13.7) 7 (9.0)
COPD 46 (61.3) 44 (95.7) 30 (66.7) 48 (64.9) 77 (96.2) <0.001
Spirometry
FEV1, L 1.65±0.51 1.46±0.46 1.51±0.53 1.72±0.58 1.53±0.39 0.070
FEV1, % 73.9±21.7 63.3±19.0 69.7±25.9 79.7±24.9 67.6±14.5 0.001
FVC, L 2.71±0.62 2.69±0.63 2.70±0.68 2.78±0.76 2.80±0.60 0.885
FVC, % 79.0±15.3 76.2±14.3 81.4±17.7 83.1±17.7 80.8±14.3 0.280
FEV1/FVC 61.3±13.8 54.3±11.7 56.2±15.0 62.6±14.7 55.2±11.3 <0.001
CAT score 17.2±6.5 15.9±6.4 17.9±8.0 16.4±7.4 16.7±7.0 0.596
Duration of current device use, mo 32.03±4.37 27.20±5.01 72.71±8.51 20.89±3.22 66.05±6.18 0.011
Duration of any inhaler use, mo 47.84±5.02 43.15±6.32 88.74±10.36 52.92±6.36 78.73±6.18 <0.001
Co-morbidity
Hypertension 35 (46.7) 22 (47.8) 20 (44.4) 45 (60.8) 36 (45.0) 0.270
DM 23 (30.7) 10 (21.7) 13 (28.9) 24 (32.4) 18 (22.5) 0.546
CVD 0 (0) 3 (6.5) 3 (6.7) 9 (12.2) 6 (7.5) 0.020
Malignancy 6 (8.0) 1 (2.2) 3 (6.6) 3 (4.1) 6 (7.5) 0.486

Values are presented as mean±standard deviation or number (%).

Kruskal-Wallis test was done for comparison of constant data. p<0.05 for two-way comparison.

BMI: body mass index; COPD: chronic obstructive pulmonary disease; FEV1: forced expiratory volume in 1 second; FVC: forced vital capacity; CAT: COPD Assessment Test; DM: diabetes mellitus; CVD: cerebrovascular disease.

Table 2.
Inhaler use assessment for five inhaler types at the first visit
Evohaler (n=75) Respimat (n=46) Turbuhaler (n=45) Ellipta (n=74) Breezhaler/Handihaler (n=80) p-value
Acceptable (no or 1 error) 28 (37.3) 22 (47.8) 28 (62.2)* 45 (60.8)* 35 (43.8) 0.026
Mean score of checklist 2.89±1.54 3.46±1.38 3.71±1.16 3.59±1.13 3.45±1.05 0.014

Values are presented as number (%) or mean±standard deviation.

The checklist for every inhaler type consists of five steps.

* There were significant differences with Evohaler (p<0.05). Total score of each checklist per patient was 5.

Fisher exact test was done for comparison of five types inhaler (p=0.026).

Kruskal-Wallis equality-of-populations rank test was done for comparison of five inhaler’ types’ scores (p=0.014).

Table 3.
Inhaler use assessment for five inhaler types at 3 months later after education
Evohaler (n=75) Respimat (n=46) Turbuhaler (n=45) Ellipta (n=74) Breezhaler/Handihaler (n=80) p-value
Acceptable (no or 1 error) 40 (64.5) 31 (81.6) 29 (76.3) 43 (66.2) 49 (65.3) 0.28*
Mean score of checklist 3.79±1.19 4.11±1.03 4.13±0.84 3.86±0.97 3.84±1.12 0.42

Values are presented as number (%) or mean±standard deviation.

The checklist for every inhaler type consists of five steps.

* Fisher exact test was done for comparison of five types of inhaler (p=0.28). Total score of each checklist per patient was 5.

Kruskal-Wallis equality-of-populations rank test was done for comparison of five inhaler types’ scores (p=0.42).

Table 4.
Factors affecting acceptable inhaler use
Variable Univariate analysis
Multivariate analysis
OR 95% CI p-value OR 95% CI p-value
Sex
  Male Reference
Female 1.440 0.564-3.681 0.446
Age 0.995 0.960-1.032 0.788
BMI 1.100 1.032-1.173 0.003 1.075 1.002-1.154 0.044
Smoking 0.365
Never smoker Reference
Ex-smoker 0.565 0.255-1.249 0.158
Current smoker 0.574 0.234-1.406 0.225
CAT 0.967 0.936-0.999 0.046 0.960 0.927-0.994 0.023
Inhaler types 0.017
Evohaler Reference Reference 0.043
Respimat 1.539 0.731-3.238 0.256 1.959 0.856-4.482 0.111
Turbuhaler 2.765 1.289-5.929 0.009 2.720 1.175-6.292 0.019
Ellipta 2.605 1.345-5.044 0.005 2.699 1.344-5.423 0.005
Handihaler/Breezhaler 1.306 0.686-2.485 0.417 1.568 0.748-3.289 0.233
Educational level 0.045 0.084
Lower (<6 yr) Reference
Middle (6-12 yr) 0.767 0.476-1.238 0.278 0.680 0.410-1.128 0.135
High (>12 yr) 2.000 0.918-4.357 0.081 1.542 0.671-3.542 0.307
Diagnosis
Asthma Reference
COPD 0.658 0.391-1.109 0.116 0.633 0.339-1.183 0.152
Comorbidities
DM 1.331 0.813-2.178 0.255
HTN 0.904 0.583-1.402 0.653
CVA 1.726 0.695-4.285 0.240
CKD 0.704 0.349-1.421 0.328
Period of current device 0.999 0.995-1.004 0.797
Total period of entire inhaler 1.003 0.999-1.007 0.209 1.004 0.999-1.009 0.126
Spirometry
FEV1, % 1.009 0.999-1.019 0.091
FVC, % 1.008 0.994-1.022 0.239
Satisfaction 0.945 0.779-1.147 0.568

Univariate and multivariate logistic regression analyses were performed to find factors associated with acceptable inhaler use in baseline assessment.

OR: odds ratio; CI: confidence interval; CAT: COPD Assessment Test; COPD: chronic obstructive pulmonary disease; DM: diabetes mellitus; HTN: hypertension; CVA: cerebrovascular accident; CKD: chronic kidney disease; FEV1: forced expiratory volume in 1 second; FVC: forced vital capacity.

Table 5.
Correct use ratio at each step with five types of inhaler device
Evohaler (n=75) Respimat (n=46) Turbuhaler (n=45) Ellipta (n=74) Breezhaler/Handihaler (n=80) p-value
Assembling
  1st visit 48 (64) 42 (91.3) 32 (71.1) 74 (100) 80 (100) <0.001
2nd visit 48 (77.4) 37 (97.4) 37 (97.4) 62 (95.4) 73 (97.3) <0.001
Holding the inhaler correctly
1st visit 71 (94.7) 41 (89.1) 45 (100) 67 (90.5) 75 (93.8) 0.166
2nd visit 59 (95.2) 37 (97.4) 37 (97.4) 61 (93.8) 66 (88.0) 0.313
Breathing out fully
1st visit 24 (32) 24 (52.2) 14 (31.1) 25 (33.8) 23 (28.8) 0.105
2nd visit 35 (56.5) 17 (44.7) 16 (42.1) 25 (38.5) 30 (40.0) 0.267
Inhalation
1st visit 36 (48.0) 28 (60.9) 42 (93.3) 61 (82.4) 66 (82.5) <0.001
2nd visit 42 (67.7) 33 (86.8) 34 (89.5) 57 (87.7) 64 (85.3) 0.023
Holding the breath
1st visit 38 (50.7) 24 (52.2) 34 (75.6) 39 (52.7) 32 (40.0) 0.005
2nd visit 50 (80.6) 31 (81.6) 32 (84.2) 45 (69.2) 54 (72.0) 0.186

Values are presented as number (%) of patients performing the device correctly to the total number of patients. Inhalation included coordination between inhalation and actuation. Fisher exact test was done. 1st visit patients n=320, 2nd visit patients n=278 (Evohaler, n=62; Respimat, n=38; Turbuhaler, n=38; Ellipta, n=65; and Breezhaler/Handihaler, n=75).

References

1. Dreher M, Muller T. Add-on therapy for symptomatic asthma despite long-acting beta-agonists/inhaled corticosteroid. Tuberc Respir Dis 2018;81:1-5.
crossref
2. Park YB, Rhee CK, Yoon HK, Oh YM, Lim SY, Lee JH, et al. Revised (2018) COPD clinical practice guideline of the Korean Academy of Tuberculosis and Respiratory Disease: a summary. Tuberc Respir Dis 2018;81:261-73.
crossref
3. Miravitlles M, Baek S, Vithlani V, Lad R. Optimal bronchodilation for COPD patients: are all long-acting β2-agonist/longacting muscarinic antagonists the same? Tuberc Respir Dis 2018;81:198-215.
crossref
4. Rogliani P, Calzetta L, Coppola A, Cavalli F, Ora J, Puxeddu E, et al. Optimizing drug delivery in COPD: the role of inhaler devices. Respir Med 2017;124:6-14.
crossref pmid
5. Bosnic-Anticevich SZ. Continued innovation in respiratory care: the importance of inhaler devices. Tuberc Respir Dis 2018;81:91-8.
crossref
6. Lindgren S, Bake B, Larsson S. Clinical consequences of inadequate inhalation technique in asthma therapy. Eur J Respir Dis 1987;70:93-8.
pmid
7. Epstein SW, Manning CP, Ashley MJ, Corey PN. Survey of the clinical use of pressurized aerosol inhalers. Can Med Assoc J 1979;120:813-6.
pmid pmc
8. Sanchis J, Gich I, Pedersen S; Aerosol Drug Management Improvement Team (ADMIT). Systematic review of errors in inhaler use: has patient technique improved over time? Chest 2016;150:394-406.
crossref pmid
9. Bosnic-Anticevich SZ. Inhaler device handling: have we really started to address the problem? Eur Respir J 2017;49:1700120.
crossref pmid
10. Molimard M, Raherison C, Lignot S, Balestra A, Lamarque S, Chartier A, et al. Chronic obstructive pulmonary disease exacerbation and inhaler device handling: real-life assessment of 2935 patients. Eur Respir J 2017;49:1601794.
crossref pmid
11. Armitage JM, Williams SJ. Inhaler technique in the elderly. Age Ageing 1988;17:275-8.
crossref pmid
12. Molimard M, Raherison C, Lignot S, Depont F, Abouelfath A, Moore N. Assessment of handling of inhaler devices in real life: an observational study in 3811 patients in primary care. J Aerosol Med 2003;16:249-54.
crossref pmid
13. Wieshammer S, Dreyhaupt J. Dry powder inhalers: which factors determine the frequency of handling errors? Respiration 2008;75:18-25.
crossref pmid
14. Mahon J, Fitzgerald A, Glanville J, Dekhuijzen R, Glatte J, Glanemann S, et al. Misuse and/or treatment delivery failure of inhalers among patients with asthma or COPD: a review and recommendations for the conduct of future research. Respir Med 2017;129:98-116.
crossref pmid
15. Chrystyn H, van der Palen J, Sharma R, Barnes N, Delafont B, Mahajan A, et al. Device errors in asthma and COPD: systematic literature review and meta-analysis. NPJ Prim Care Respir Med 2017;27:22.
crossref pmid pmc
16. Kerwin EM, Preece A, Brintziki D, Collison KA, Sharma R. ELLIPTA dry powder versus metered-dose inhalers in an optimized clinical trial setting. J Allergy Clin Immunol Pract 2019;7:1843-9.
crossref pmid
17. Collison KA, Patel P, Preece AF, Stanford RH, Sharma RK, Feldman G. A randomized clinical trial comparing the ELLIPTA and HandiHaler dry powder inhalers in patients with COPD: inhaler-specific attributes and overall patient preference. COPD 2018;15:46-50.
crossref pmid
18. Liang CY, Chen YJ, Sheu SM, Tsai CF, Chen W. Misuse of inhalers among COPD patients in a community hospital in Taiwan. Int J Chron Obstruct Pulmon Dis 2018;13:1309-16.
crossref pmid pmc
19. Lee H, Boo S, Lim Y, Kim S, Kim IA. Accuracy of inhaler use in patients with chronic obstructive pulmonary disease. Clin Nurs Res 2014;23:560-74.
crossref pmid
20. Vanderman AJ, Moss JM, Bailey JC, Melnyk SD, Brown JN. Inhaler misuse in an older adult population. Consult Pharm 2015;30:92-100.
crossref pmid
21. Duarte-de-Araujo A, Teixeira P, Hespanhol V, Correia-de-Sousa J. COPD: misuse of inhaler devices in clinical practice. Int J Chron Obstruct Pulmon Dis 2019;14:1209-17.
pmid pmc
22. Melani AS, Bonavia M, Mastropasqua E, Zanforlin A, Lodi M, Martucci P, et al. Time required to rectify inhaler errors among experienced subjects with faulty technique. Respir Care 2017;62:409-14.
crossref pmid
23. Altman P, Bergna MA, Garcia GR, Guerin T, Pino AV, Whiteford JL. Patient perception of Breezhaler((R)) and Ellipta((R)) device feedback mechanisms in COPD: the ADVANTAGE study. Curr Med Res Opin 2019;35:221-7.
pmid
24. Svedsater H, Dale P, Garrill K, Walker R, Woepse MW. Qualitative assessment of attributes and ease of use of the ELLIPTA dry powder inhaler for delivery of maintenance therapy for asthma and COPD. BMC Pulm Med 2013;13:72.
crossref pmid pmc
25. Klijn SL, Hiligsmann M, Evers S, Roman-Rodriguez M, van der Molen T, van Boven JF. Effectiveness and success factors of educational inhaler technique interventions in asthma and COPD patients: a systematic review. NPJ Prim Care Respir Med 2017;27:24.
pmid pmc
26. Melani AS, Zanchetta D, Barbato N, Sestini P, Cinti C, Canessa PA, et al. Inhalation technique and variables associated with misuse of conventional metered-dose inhalers and newer dry powder inhalers in experienced adults. Ann Allergy Asthma Immunol 2004;93:439-46.
crossref pmid
27. Rydman RJ, Sonenthal K, Tadimeti L, Butki N, McDermott MF. Evaluating the outcome of two teaching methods of breath actuated inhaler in an inner city asthma clinic. J Med Syst 1999;23:349-56.
pmid
28. Basheti IA, Armour CL, Bosnic-Anticevich SZ, Reddel HK. Evaluation of a novel educational strategy, including inhaler-based reminder labels, to improve asthma inhaler technique. Patient Educ Couns 2008;72:26-33.
crossref pmid
29. Geller DE. Comparing clinical features of the nebulizer, metered-dose inhaler, and dry powder inhaler. Respir Care 2005;50:1313-21.
pmid


ABOUT
ARTICLE & TOPICS
Article category

Browse all articles >

Topics

Browse all articles >

BROWSE ARTICLES
FOR CONTRIBUTORS
Editorial Office
101-605, 58, Banpo-daero, Seocho-gu (Seocho-dong, Seocho Art-Xi), Seoul 06652, Korea
Tel: +82-2-575-3825, +82-2-576-5347    Fax: +82-2-572-6683    E-mail: katrdsubmit@lungkorea.org                

Copyright © 2024 by The Korean Academy of Tuberculosis and Respiratory Diseases. All rights reserved.

Developed in M2PI

Close layer
prev next