The Xpert MTB/RIF assay (Xpert assay; Cepheid, Sunnyvale, CA, USA) is a fully automated, real-time polymerase chain reaction assay that is designed to detect the presence of
Mycobacterium tuberculosis (MTB) and rifampin (RFP) resistance
1. The Xpert assay can be utilized as a rapid diagnostic test for tuberculosis and RFP-resistant tuberculosis
1. The review article which was recently published in this journal emphasized rapid molecular diagnosis of RFP-resistant tuberculosis for the optimal treatment of multidrug-resistant tuberculsosis
2.
Xpert assay can also be used to differentiate MTB from nontuberculous mycobacteria (NTM) in acid-fast bacilli (AFB) smear-positive cases
3. Although several previous studies have demonstrated that the Xpert assay showed excellent performance in distinguishing MTB from NTM in smear-positive or negative clinical specimens
1,4, the misdiagnosis of MTB by the Xpert assay was recently reported for five NTM species (
M. abscessus,
M. marinum,
M. smegmatis,
M. phlei, and
M. aurum) at a high bacterial load
5. The results reported by Pang et al.
5 have very important clinical implications, because AFB smear microscopy cannot differentiate between MTB and NTM and many patients with NTM pulmonary disease were misdiagnosed with pulmonary tuberculosis and even multidrug-resistant tuberculosis
6. Moreover,
M. abscessus is an important etiologic organism of NTM pulmonary disease in many countries with high tuberculosis burden such as China and South Korea
7,8. Misdiagnosis of NTM as MTB or RFP-resistant MTB is a significant concern in defining a proper tuberculosis program in these countries. We therefore tried to validate the previous experiment with the misdiagnosed NTM species to evaluate whether the Xpert assay showed cross-reactivity between MTB and five NTM species.
A total of five NTM reference strains (
M. abscessus,
M. marinum,
M. smegmatis,
M. phlei, and
M. aurum) from American Type Culture Collection or the Korean Mycobacteria Resource Center were included in this study (
Table 1). Freshly grown bacterial colonies were obtained from 3% Ogawa agar (Shinyang, Seoul, Korea). Subsequently, the bacteria were suspended in sterile distilled water and vortexed vigorously. The turbidity was adjusted to a 0.5 McFarland (approximately 1×10
7 colony-forming unit [CFU]/mL)
9. Additionally, the bacteria were diluted, resulting in tubes containing 1×10
5 CFU/mL equivalents. The Xpert assay was performed using the G4 version of cartridges with sediment protocol according to the manufacturer's instructions.
The Xpert results for different NTM species at two concentrations are summarized in
Table 1. All NTM species gave negative results, and non-specific positive reactions were not observed in any probes: no changes in threshold cycle value were observed. The sample processing controls in all tested bacterial suspensions met the acceptance criteria, and no invalid results were observed for any sample.
The manufacturer presented the data for analytical specificity and exclusivity in the package insert
10. A total of 24 NTM species, including
M. abscessus,
M. marinum, and
M. smegmatis were tested at 10
8 CFU/mL or 10
6 CFU/mL. No MTB results were generated from NTM strains except
M. scrofulaceum at 10
8 CFU/mL (one of three replicates). Moreover, numerous studies provided data on a variety of NTM that grew from the specimens tested to look for evidence of cross-reactivity
1. A systematic review on the diagnostic accuracy of the Xpert assay reported that only one specimen (0.6%) exhibited a false-positive result among 180 specimens that grew NTM
1. On the other hand, Pang et al.
5 reported that misdiagnosis of MTB, even RFP-resistant MTB by the Xpert assay was observed in five NTM species at a high bacterial load, although these results were obtained on the basis of artificial non-human samples. They regarded this cross-reactivity as an intrinsic problem of the Xpert assay. A high burden of organisms might cause fluorescence cross-talk in the detection channels, which could lead to false-positive results. However, other possible causes of false-positivity should be excluded such as specimen contamination at the time of sample preparation, since an artificial bacterial suspension was used. Indeed, we could not reproduce their findings in our study: we did not observe any false-positive reactions in the five NTM species, even at a higher concentration of 1×10
7 CFU/mL equivalents. In light of these data, it is difficult to interpret the false-positive results for five NTM species at a high bacterial load as a systematic cross-reactivity issue in the Xpert assay. However, our study has limitations. Since only one strain per species was used in this study, we cannot exclude the possibility of cross-reactivity with some NTM strains.
In conclusion, we tested the Xpert assay for cross-reactivity between MTB and five NTM species at high bacterial load and did not observe any false-positive reactions. Therefore, the previously reported cross-reactivity between MTB and five NTM species may not be a systematic problem with the Xpert assay.
Table 1
Xpert assay results for detecting different mycobacterial species
Species |
Strain(s) and source |
Xpert result at different bacterial loads (CFU/mL equivalent) |
107
|
105
|
M. abscessus
|
ATCC19977 |
ND |
ND |
M. aurum
|
ATCC51345 |
ND |
ND |
M. marinum
|
KMRC 00136-21108 |
ND |
ND |
M. phlei
|
ATCC11758 |
ND |
ND |
M. smegmatis
|
ATCC19420 |
ND |
ND |